Features:
601044 added in the battery back-up system,other functions just like 601043 system . The tourniquet is designed to be operated primarily from an electrical outlet connection. Internal rechargeable batteries are provided as a means of powering the tourniquet when an electrical outlet is not available or in the event of power failure.
Functions:
Faster inflation and time-lapse deflation;
The micro processor monitors and gives alarm kink elapsed time;
Electronically precise pressure setting;
Automatic compensation of small leakage;
Pressure no more than 600mmHg;
Computerized memory;
Backup mode for pressure sustaining in case of power failure and patient transferring;
Battery back-up system automatically engages if AC current interrupted.
Specificationsons:
Cuff pressure: Adjustable 0 to 600mmHg with 5mmHg resolution;
Inaccuracy: +/-5mmHg or +/-2% of reading which ever is greatest;
Time setting: Adjustable 0 to 120 minutes with 1 min resolution;
Line power: 220V AC(±10%), 50Hz;
Initial flushes time: ≤60s.